Principles
|
: |
Clotting
: Scattered Light Detection Method.
Percentage Test End Point Detection Method.
Chromogenics * : Colorimetric Method.
|
|
Quality Control
|
: |
X
control, Levy-Jennings Control.
12 files, 7 parameters, 180 data points in memory. |
|
Parameters
|
: |
PT
(Prothrombin Time).
APTT (Activated Partial Thromboplastin Time).
Fbg (from Clauss-Method).
TTO (Thrombotest).
Normotest.
TT (Thrombin Time).
AT - III.
|
|
Data Storage
|
: |
300
samples (1500 tests) results with the latest 600 tests of
coagulation curves. |
|
Programmable Test Protocol
|
: |
4
Channels.
The operator can program the test protocol.
|
|
Print Out
|
: |
Built-in
thermal graphic printer. |
|
Throughput
|
: |
Max.
: Approx. 50 tests per hour.
Average (PT, APTT, Fbg) Approx. 33 tests per hour.
|
|
Dimensions
W x H x D (mm)
|
: |
Main
unit : Approx. 540 x 487 x 470 |
|
Cycle Time
|
: |
PT,
Fbg : Approx. 100 seconds.
Others : Approx. 190 seconds.
Automatic Extension : Up to 600 seconds.
|
|
Weight (kg)
|
: |
Main
unit : Approx. 45. |
|
Sample Volume
|
: |
PT
: 50µL.
APTT : 50µL.
Fbg : 10µL.
Others: Various.
|
|
Power Requirement
|
: |
117
/ 230V AC +/- 15% (50 or 60Hz). |
|
Reaction Tubes
|
: |
Capacity
: 60 tubes.
|
|
Power Consumption
|
: |
Less
than 400VA (50 or 60Hz). |
|
Sample Aspiration
|
: |
The
sample probe, with a liquid surface sensor, quantitatively
aspirates plasma from centrifuged whole blood samples.
|
|
|
|
|
|
* Only for CA-539/540
|
| |
|
|
|
|
|