The EASY BLOOD GAS ANALYZER

Easy Blood Gas Analyser

 Blood Gas Analyzer operation has never been simpler

Unique electrode design, combined with precise control of calibrator volumes, ensuring economical operation and low cost per sample.

 

 

  • Compact reagent module for convenience, economy and safety.

  • Disposable, maintenance-free electrodes.

  • EasyBloodGas can be maintained by anyone, anytime, anywhere.

  • Comprehensive quality control and data management.

  • Simple menu-prompted operation with a touch of the keypad.

  

TECHNICAL SPECIFICATIONS
           

CLIA Classification

: Moderate complexity.  

Ambient Conditions

: 15 - 30°C (59-86°F), 500 - 800 mmHg (max 15 PSI),
5 - 85% relative humidity, non-condensing,
atmospheric air environment (21% PO2).
 

Sample Type

: Whole blood.  

Analysis Time

: 125 seconds.

    

Sample Size

: 100 µL Syringe 75 µL Capillary.  

Data Storage

: 64 Patient results with Operator ID, Patient ID, Data and Time .
QC - up to 30 results for each Level (1, 2, 3).
 

Measured Parameters

: pH      : 6.900 - 7.900 pH units.
PCO2 : 8.0 - 150.0 mmHg.
PO2   : 10 - 700 mmHg.
 
 

Calibration

: Automatic or On-Demand.

Calculated Parameters

: pH (T) 
(pH temperature corrected).

PCO2 (T) 
(PCO2 temperature corrected).

PO2 (T) 
(PO2 temperature corrected).

TCO2   :  0 - 50 mmol/L.
(Total Carbon dioxide) 

HCO3-  : 0 - 50 mmol/L.
(Bicarbonate)

BEb      :  -25.0 to 25.0 mmol/L.
(Base Excess in blood)

BEecf   : -25.0 to 25.0 mmol/L.
(Base Excess in extra cellular fluid) 

SBC      : 0 - 50 mmol/L.
(Standard Bicarbonate)

%SO2c : 40.0 - 100.0%.
(Oxygen Saturation calculated at normal P50)

A-aDO2 : 0 - 700 mmHg.
(Alveolar arterial oxygen gradient)

RI           : 0.0 - 70.0.
(Respiratory Index) 
 
 

Input/Output

: Numeric keypad, graphic display, 27 column thermal printer, barcode reader port, RS-232 computer interface port.
Input Parameters : Patient Temperature : 5 - 45°C.
Hemoglobin               : 3.0 - 30.0 g/dL.
FIO2                          : 10 - 100%.
Patient ID                   : 14 digits.
 
  Power : 100/115~VAC, 50-60 Hz, 0.8 A or 220~VAC, 50-60 Hz, 0.4 A.
Sample Temperature Control : 37°C + 0.2°C.   Size & Weight : 14.5"W x 12.5"H x 7"D (37cm W x 32m H x 18cm D),
16lbs (7.3kg) with reagent module.
         